9 Installation
If the device mains voltage needs to be changed, change the fuses accordingly. See
chapter
Changing the fuses
for information on how to install the fuses.
9.9.3 Changing the fuses
Main fuses are located at the rear of the carriage, under the back panel cover. To reach
the fuses, remove the cover first.
The device is supplied with pre-installed 10A fuses. The fuses may need to be replaced
as part of the installation depending on the mains voltage:
• 10A suited for 220-240 VAC mains voltage
• 15A suited for 100-120 VAC mains voltage
To release a fuse cap, push it inwards and rotate counterclockwise with a flat head
screwdriver.
Remove the fuse from the socket and replace it with a new one. Install the fuse by
pushing the fuse cap in and rotating it clockwise with a screwdriver.
The 10A and 15A fuses are physically of different sizes and have their own fuse caps. If
you need to change the fuse rating, also the fuse cap needs to be changed. The fuses
are supplied with the corresponding fuse caps.
CAUTION!
Use only specified fuses. See the list of fuses from chapter
Technical
specifications
.
9.9.4 Additional wall bracket installation
If the device cannot be bolted to the floor and the exhibition stand is not used, an
additional wall bracket is required (ordered separately). This part of the installation can
be done after the column has been unpacked.
NOTICE!
The existing wall bracket holder doesn't need to be removed when
installing an additional holder.
86
ORTHOPANTOMOGRAPH
™
OP 3D