9 Installation
9.3 Back panel connectors
1.
Power cord
2.
Ethernet cable for workstation connection
3.
Exposure button connector
4.
Remote exposure switch connector (optional)
5.
External warning light connector (optional)
6.
Main fuses F1+F2 (10A fuses installed as default)
7.
External warning light (optional) fuse F3
8.
Control interface
NOTICE!
All cables, except the power cord, must be routed through a strain relief
attached to the back panel plate.
The Control interface can be used to connect optional device controls such as external
exposure switch and door switch.
PIN
1
2
3
4
5
6
7
8
SIGNAL
EXPSW
READY
GND +5V
WARNING
DOORSW
USE
Exposure switch
connection
LOW signal when
device is ready
LOW signal when
exposure is ON
Door switch
connection
NOTICE!
If a door switch is installed it has to be activated on the GUI. See
chapter
Device configuration
for more information.
9.4 Installing the device
NOTICE!
If the device cannot be fixed to the floor or the device is to be installed
on an exhibition stand, see chapter
Additional installation procedures
for more
details.
1.
Transport the packing pallet to the installation location.
2.
Remove the packing straps.
3.
Remove the package lid and sides.
4.
Move the boxes with accessories and device covers aside.
ORTHOPANTOMOGRAPH
™
OP 3D
57