9 Installation
9.8 Complete device calibration
After the installation of the device and during the annual maintenance, a complete
calibration procedure must be performed. See chapter
Preparation for Calibration
for
information on the unit calibration.
Perform the calibrations in the order as described in this chapter.
After the calibrations have been completed, the QC images need to be taken and stored
for the user as reference.
9.8.1 Accessing the calibrations
A complete list of calibration programs, required to be performed after the unit
installation, can be found in the PIN code protected Service menu.
NOTICE!
The calibrations programs in the Service menu are intended for service
technicians only, and not to be performed by the end user of the unit.
NOTICE!
The device's country selection needs to be configured correctly before
performing the calibrations. See chapter
Device configuration
.
1.
Turn the device on and establish a connection to the workstation. See chapters
Powering on the device
and
Preparing the device for imaging
for more information.
2.
Select
Settings
menu on the GUI.
3.
Select
Service menu
and enter PIN code
0612
.
4.
The GUI informs that additional options have been added under the Service menu.
5.
Select
Calibrations
under the Service menu to see the list of all calibration
programs.
ORTHOPANTOMOGRAPH
™
OP 3D
73