![KaVo ORTHOPANTOMOGRAPH OP 3D User And Installation Manual Download Page 83](http://html2.mh-extra.com/html/kavo/orthopantomograph-op-3d/orthopantomograph-op-3d_user-and-installation-manual_3750531083.webp)
9 Installation
9.8.12 Panoramic beam size verification (optional)
This program runs an automatic test to verify, that the X-ray beam hits the active area of
the sensor with the required accuracy.
NOTICE!
The Panoramic Beam Size Verification is an optional, yet
recommendable, procedure and is mandatory to be performed where local
regulations require it.
1. GUI:
Go to Settings menu.
2. GUI:
Select Quality Control menu.
3. GUI:
Select
Panoramic Beam Size Verification
program from the list.
NOTICE!
The Panoramic Beam Size Verification program appears under
Quality Control menu only when the Service menu has been unlocked.
4.
Protect yourself from radiation.
5.
Press and hold down the exposure switch to take the calibration.
6.
The test result appears on the GUI.
7.
Acknowledge the result by pressing the
OK
button on the GUI.
ORTHOPANTOMOGRAPH
™
OP 3D
83