9 Installation
36.
Connect the Mains cable (
1
), Ethernet cable (
2
) and Exposure button cable (
3
) to
the rear of the carriage.
NOTICE!
Route the Ethernet cable and exposure button cable through the
strain relief.
NOTICE!
You can also connect other additional cables to the carriage at this
point. See chapter
Back panel connectors
.
NOTICE!
The cables will move when the unit is driven upwards. Ensure that
the cables connected to the device are long enough so that the unit can be
driven to upmost position without causing any strain to the cables.
9.5 Attaching the covers
NOTICE!
The screws used to attach the covers are located in the screw bag.
1.
Attach upper shelf covers. Place the right cover onto the upper shelf and then push
the left cover against it.
2.
Secure the covers with two screws on the top of the covers and then one on the
bottom.
ORTHOPANTOMOGRAPH
™
OP 3D