Starting and Configuring Chemical
Purge
Using Buttons on Motherboard
Start manual chemical purge by pressing simultaneously two
PURGE
buttons on the motherboard inside the transmitter for a few seconds.
Indicator led flashes until purge is complete (up to 6 minutes).
0508-011
Figure 63
Purge Buttons on Motherboard
Using Display/Keypad (Optional)
Set the automatic and manual chemical purge by using the
display/keypad.
1.
Open the
MAIN MENU
by pressing any of the ▼▲◄► arrow
buttons.
2.
Select ►
Measuring
, press ►button
.
3.
Select ►
Chemical purge
, press ►button.
0706-006
Figure 64