User's Guide _______________________________________________________________________
40 _______________________________________________________________________________
0503-033
Figure 24
Cleaning of Tightening Cone
The following numbers refer to Figure 24:
1
=
Fitting screw
2
=
Sealing washer
3
=
Tightening cone
4
=
Clean cotton stick
CAUTION
In pressurized processes it is essential to tighten the supporting nuts
and screws very carefully to prevent loosening of the probe by the
action of pressure.
NOTE
When installed in a process with a pressure differing from normal
atmospheric pressure, please enter the pressure value of the process
(in hPa or mbar) into the transmitter memory via the serial line (see
command PRES and XPRES on page 94) or by using
display/keypad.
907023/335 for High Temperatures
907023/335 is installed similarly than the 907023/333 probe but
without the supporting bar. Refer to Appendix A on page 163 for
more information on the duct installation kit for 907023/335.
To avoid incorrect humidity readings, the temperature differences
between inside and outside of the duct must not be remarkable.