User's Guide _______________________________________________________________________
96 _______________________________________________________________________________
User Port Serial Settings
Using Display/Keypad
The communication settings for the user port can be changed via the
serial line or by using the optional display/keypad. The
communication settings for the service port are fixed and not
changeable.
1.
Press any of the arrow buttons to open the MAIN MENU.
2.
Select
Interfaces
and press the ►arrow button to confirm your
selection.
3.
Select
Serial interface
and press the ►arrow button to confirm
your selection.
4.
Select
Bit rate/Serial format/Comm. mode
by pressing the
CHANGE
button. Use the
▲▼
arrow buttons to select and
press
SELECT
to confirm your selection.
5.
If you selected RUN communication mode, select
RUN
interval
for and press
SET
to confirm your selection.
6.
Use the arrow buttons to set the measuring interval and the unit.
Press
OK
to confirm your settings.
7.
If you selected POLL communication mode, select POLL
address and press
SET
to confirm your selection.
8.
Use the arrow buttons to set the transmitter address. Press
OK
to
confirm the setting.
9.
Use the arrow buttons to select
ECHO
. Press
ON
to turn to it
on. Press
OFF
to turn it off.
10. Press
EXIT
to return to the basic display.
The new user port settings set using the display/keypad are effective
immediately.