Chapter 7 _____________________________________________________________ Technical Data
JUMO _________________________________________________________________________ 147
CHAPTER 7
TECHNICAL DATA
This chapter provides the technical data of the product.
Specifications
Performance
Relative Humidity
Measurement range
0 ... 100 %RH
Accuracy (including non-linearity, hysteresis and repeatability)
with HUMICAP
®
180
for typical applications
HUMICAP
®
180R
for typical applications (standard)
HUMICAP
®
180C
for applications with chemical purge
and/or warmed probe
HUMICAP
®
180RC
for applications with chemical purge
and/or warmed probe
at +15 ... 25 °C
± 1 % RH (0 ... 90 % RH)
± 1.7 % RH (90 ... 100 %RH)
at -20 ... +40 °C
± (1.0 + 0.008 × reading) % RH
at -40 ... + 180 °C
± (1.5 + 0.015 × reading) % RH
with HUMICAP
®
180L2
for applications with demanding chemical
environment
at -10 ... +40
°
C
±
(1.0 + 0.01 × reading) % RH
at -40 ... +180
°
C
±
(1.5 + 0.02 × reading) % RH
Factory calibration uncertainty (+20 °C)
±0.6 % RH (0 ... 40 % RH)
±1.0 % RH (40 ... 97 % RH)
(Defined as ± 2 standard deviation limits.
Small variations possible, see also
calibration certificate.)