7.
Select
Scale,
lower limit,
by pressing the ▲▼arrow buttons.
Press
SET
to confirm your selection. Press
OK
to confirm your
setting.
8.
Select the upper limit
by pressing the ▲▼arrow buttons. Use
the arrow buttons to set the upper limit value. Press
SET
to
confirm your selection. Press
OK
to confirm your setting.
9.
Press
EXIT
to return to the basic display.
AMODE/ASEL
Use the serial line to select and scale the analog output quantities.
Connect the transmitter to the PC. Open the terminal connection
between your PC and the transmitter.
1.
Check the analog output modes with the
AMODE
command.
Example:
>amode
Ch1 output : 0...1V
Ch2 output : 0...1V
>
2.
Select and scale the quantities for the analog outputs with the
command
ASEL
. Note that the optional quantities can be
selected only if they have been selected when ordering the
device.
ASEL
[
xxx yyy zzz
]
where
xxx
=
Quantity of channel 1
yyy
=
Quantity of channel 2
zzz
=
Quantity of the optional analog output channel 3
Enter always all the quantities for all outputs. For quantities and their
abbreviations see Table 1 on page 16 and Table 2 on page 17.
Use the command
ASEL
[
xxx yyy
] as shown in the example below
when using a device with two analog outputs.
Example:
>asel rh t
Ch1 (RH ) low : 0.00 %RH ? 0
Ch1 (RH ) high : 100.00 %RH ? 100
Ch2 (T ) low : -40.00 'C ? -50
Ch2 (T ) high : 60.00 'C ? 80
>