Day-to-Day Operation
4
STERRAD
®
200 User’s Guide
57
Note:
The exact sequence of cancellation actions depends on
which process was in operation at the time of cancellation.
7. The display indicates when cancellation is complete and a summary display
appears.
Loads from canceled cycles should be re-wrapped using new polypropylene
wrap, STERRAD Chemical Indicator Strips, and STERRAD SealSure
®
Chemical
Indicator Tape. If a biological indicator was used in the canceled load, a new one
should be placed in the chamber before restarting the new cycle.
WARNING! IF A CYCLE CANCELS AND THE ITEMS IN THE
LOAD APPEAR WET, HYDROGEN PEROXIDE MAY
BE PRESENT. WEAR CHEMICAL RESISTANT
LATEX, PVC (VINYL), OR NITRILE GLOVES WHILE
REMOVING THE ITEMS FROM THE CHAMBER,
AND TO WIPE OFF THE ITEMS WITH A DAMP
CLOTH. DISPOSE OF THE CONTAMINATED CLOTH
ACCORDING TO YOUR FACILITIES PROCEDURES.
Automatic Cancellation
If the sterilizer control system cancels a cycle, the display indicates when the
cancellation process is complete. As with manual cancellation (above), the load
should be repackaged using new polypropylene wraps, chemical indicators, etc.
Note the messages on the display and on the paper printout, and refer to “Chapter
6. Troubleshooting” for more information.