Day-to-Day Operation
4
STERRAD
®
200 User’s Guide
43
Figure 12. If skin contact occurs,
immediately flush the area with water.
WARNING! DO NOT REMOVE THE PLASTIC WRAPPER FROM
THE CASSETTE PACKAGE IF THE INDICATOR
STRIP IS RED. RED INDICATES DAMAGE. CALL
THE ASP CUSTOMER CARE CENTER FOR CREDIT.
To insert a cassette, do the following:
1. Confirm that the sterilizer display indicates that a new cassette is needed.
2. Confirm that the chemical indicator strip on the cassette sleeve is NOT red;
red indicates a damaged cassette and concentrated hydrogen peroxide may be
present.
3. Confirm that the cassette expiration date has not passed.
Note:
The system considers the cassette expired 8 days after
insertion regardless of the printed expiration date. The
cassette is ejected at that time. Once the cassette has been
removed, even if it was not used, the barcode is rendered
unreadable and the cassette cannot be reinserted.
Summary of Contents for ASP STERRAD 200
Page 1: ...STERRAD 200 Sterilization System User s Guide REF 99712 99712_01 May 2008...
Page 16: ......
Page 28: ......
Page 37: ...Day to Day Operation 4 STERRAD 200 User s Guide 35 STERRAD CycleSure Biological Indicator...
Page 63: ...Day to Day Operation 4 STERRAD 200 User s Guide 61 Cycle Completion Flow Chart...
Page 74: ......
Page 82: ......