Day-to-Day Operation
4
STERRAD
®
200 User’s Guide
41
3. Gently push the carriage into the sterilizer so that the foot mechanism on the
bottom of the sterilizer contacts and locks into the flange at the bottom of the
carriage. The top front of the carriage rests just inside the chamber.
4. Release the two-tier shelf using the release. Gently push the two-tier shelf
forward into the sterilizer. Make sure the alignment wheels are correctly
positioned on the rails for smooth movement. Continue pushing until the
two-tier shelf is completely inside the chamber and it engages the inside stop.
Figure 11. Push the two-tier shelf completely into the chamber.
5. Step on the carriage release and move the carriage to the desired location.
6. Verify that no part of the two-tier shelf or its contents comes in contact with
the door or electrode. Close the door and begin your desired cycle or go to
the section on “Inserting a Cassette.”