1
More Information
200 mm
250 mm
310 mm
400 mm
0 mm
125 mm
Measurements are approximate and are for reference only.
How To Determine What Can Be Sterilized In The STERRAD 200 Sterilizer
®
How To Determine What Can Be Sterilized In The STERRAD 200 Sterilizer
®
Is The Reprocessable Medical Device Made Of The Following Materials?**
Does The Reprocessable Medical Device Have A Lumen?
Lumen Lengths
May have limited life after repeated sterilization.
Contact ASP or the device manufacturer to determine the compatibility of any device manufactured from PMMA.
Delrin , Nylon , and Teflon are registered trademarks of the DuPont Corporation.
®
®
®
ULTEM Polymers is a registered trademark of the GE Company.
®
KRATON™ Polymers is a trademark of KRATON Polymers U.S. L.L.C.
- Aluminum
- Brass
- Delrin acetal resin (polyacetal)
- Ethylvinyl acetate (EVA)
- Glass
®
- KRATON
Polymers
- Neoprene
- Non-mated Nylon (polyamide)
- Polycarbonate
- Polyethylene
TM
®
- Polyetherimide (ULTEM Polymers)
- Polymethyl methacrylate (PMMA)
- Polyphenylene sulfone (Radel )
- Polypropylene
- Polystyrene
®
®
Proceed with
Processing.
Yes
Yes
Proceed With Processing If The Lumen Conforms To The Dimensions Listed Below
Is The Lumen Made Of Stainless Steel, Or Teflon ?
®
- Polyurethane
- Polyvinyl chloride (PVC)
- Silicone elastomers
- Stainless steel
- Teflon (polytetrafluoroethylene)
- Titanium
®
No/Don’t
Know
No
No/Don’t
Know
Yes
2
3
4
If the lumens do not conform to these dimensions, please call the device
manufacturer for information on how to properly sterilize these devices.
Lumens not conforming to these dimensions should not be processed in
the STERRAD 200 Sterilizer.
Please call the device manufacturer for
information on how to properly sterilize
this device.
Please call the device manufacturer for
information on how to properly sterilize
this device.
4 mm (0.157 inches)
5 mm (0.196 inches)
6 mm (0.236 inches)
3 mm (0.118 inches)
2 mm (0.078 inches)
1 mm (0.039 inches)
Inside Lumen
Diameters
More Information
* Validation testing for this lumen size was conducted using a maximum of 12 lumens per load.
Your loads should not exceed the maximum number of lumens validated by this testing.
** This list of materials does not apply to trays and containers or other packaging materials. Please refer to the STERRAD 200 User’s Guide for information on appropriate packaging
materials for use in the STERRAD System.
®
Inside Diameter
Length
Stainless Steel Lumen
1 mm or larger
2 mm or larger
3 mm or larger
125 mm or shorter*
250 mm or shorter*
400 mm or shorter
Inside Diameter
6 mm or larger
Length
Teflon /Polyethylene Lumens
®
__
310 mm or shorter
Summary of Contents for ASP STERRAD 200
Page 1: ...STERRAD 200 Sterilization System User s Guide REF 99712 99712_01 May 2008...