Provide at least 25 mm (1 inch) of space between the ceiling of the electrode
and the top of the load.
♦
Place STERRAD CycleSure
®
Biological Indicator (BI) or an ASP-approved
biological indicator in the chamber. Frequency of biological testing should
be at least once per day or in accordance with your facility’s policy. Review
the instructions included with the biological indicator to ensure proper use.
♦
Proceed to “Chapter 4. Day-to-Day Operation” for information on starting a
cycle.