For Your Safety
2
STERRAD
®
200 User’s Guide
15
Safe Maintenance
•
WARNING! HYDROGEN PEROXIDE MAY BE PRESENT
If white residue is visible on the load; this may be residue from the hydrogen
peroxide stabilizer. Wear chemical resistant latex, PVC (vinyl), or nitrile
gloves when removing a load with visible white residue. White residue can
be minimized by making sure regular Planned Maintenance (PM) procedures
are performed on your sterilizer. The sterilizer will inform you when Planned
Maintenance is due. Please schedule your PM service in a timely manner.
•
Repairs and adjustments should only be attempted by experienced
technicians who are fully trained to maintain and repair the STERRAD 200
Sterilizer.
•
Use of unauthorized parts for maintenance or repair could cause personal
injury, result in costly damage or unit malfunction, and void the warranty.
•
Do not clean the chamber door area with abrasives. The sterilization chamber
uses an O-ring vacuum seal to maintain a vacuum in the chamber. Never use
rough cleaning tools, such as a wire brush or steel wool, on the door housing
or chamber assembly. This could damage the seal.
Additional Information
•
The information contained in this chapter is repeated where appropriate
throughout this guide for your safety and use. This information is
subsequently labeled: WARNINGS, Cautions or Notes as appropriate.
•
WARNINGS
are shown in the text in all bold upper case letters. They
indicate events or conditions that can result in serious injury or death.
•
C
autions are shown in the text in bold letters. They indicate events or
conditions that can result in damage to equipment.
•
Notes are shown in the text with a check mark
. They highlight specific
information about the proper use and maintenance of the STERRAD 200
Sterilizer.
Summary of Contents for ASP STERRAD 200
Page 1: ...STERRAD 200 Sterilization System User s Guide REF 99712 99712_01 May 2008...
Page 16: ......
Page 28: ......
Page 37: ...Day to Day Operation 4 STERRAD 200 User s Guide 35 STERRAD CycleSure Biological Indicator...
Page 63: ...Day to Day Operation 4 STERRAD 200 User s Guide 61 Cycle Completion Flow Chart...
Page 74: ......
Page 82: ......