Preparing Items To Be Sterilized
18
STERRAD
®
200 User’s Guide
»
an inside diameter of 1 mm or larger and a length of 125 mm or shorter,*
»
an inside diameter of 2 mm or larger and a length of 250 mm or shorter,*
»
an inside diameter of 3 mm or larger and a length of 400 mm or shorter.
can be processed in the STERRAD 200 Sterilizer.
*Validation testing for this lumen size was conducted using a maximum of 12
lumens per load. Your loads should not exceed the maximum number of lumens
validated by this testing.
Note:
The validation studies for this sterilizer were performed
using the STERRAD cycle consisting of a 6-minute, 45-
second injection phase followed by a 2-minute diffusion
phase, and a 2-minute plasma phase. The validation
studies were performed using a validation load consisting
of four instrument trays, each weighing 9.12 lbs. (the total
weight of the load was 36.48 lbs.).
Note
: Determination of true length is based upon actual
instruments identified as intended for reprocessing and
meet all other criteria for processing in the STERRAD 200
Sterilizer.
WARNING! DO NOT ATTEMPT TO STERILIZE ITEMS OR
MATERIALS THAT DO NOT COMPLY WITH THE
GUIDELINES SPECIFIED IN THIS GUIDE. IN
ADDITION, YOU SHOULD READ THE MEDICAL
DEVICE MANUFACTURER’S INSTRUCTIONS, OR
CALL THE ASP CUSTOMER CARE CENTER TO
DETERMINE WHETHER AN ITEM CAN BE
STERILIZED BY THIS UNIT.
Summary of Contents for ASP STERRAD 200
Page 1: ...STERRAD 200 Sterilization System User s Guide REF 99712 99712_01 May 2008...
Page 16: ......
Page 28: ......
Page 37: ...Day to Day Operation 4 STERRAD 200 User s Guide 35 STERRAD CycleSure Biological Indicator...
Page 63: ...Day to Day Operation 4 STERRAD 200 User s Guide 61 Cycle Completion Flow Chart...
Page 74: ......
Page 82: ......