A
Specifications
82
STERRAD
®
200 User’s Guide
Inst
a
llation (please call ASP at 1-888-STERRAD before re-installing the sterilizer)
Space Requirements
225.4 cm H x 291.5 cm W x 213.0 cm D (88.7 in x 114.8 in
x 83.7 in)
Mobility
Freestanding unit: Movable
In-Wall Unit: Mounted movable freestanding unit or
mounted in-wall unit
Venting Requirements
None required
Gas Tank Requirements
None required
Electrical Requirements
3 m/10 ft. power cord
Weight
475 kg/1047 lbs.
Equipment Rating
High Voltage
200V~, 3~, 50/60 Hz, 4 wire, 15 A
208V~, 3~, 50/60 Hz, 5 wire, 15 A
380V~/400V~/415V~, 3~, 50/60 Hz, 5 wire, 15 A
Electrical Requirements
The STERRAD
®
200 Sterilizer should only be plugged into outlets approved by a qualified technician.
For further requirements, refer to the label on the back panel of the sterilizer or call your ASP service
representative. Only a qualified technician can determine when the STERRAD 200 Sterilizer can be
safely moved to a new power source.
Protection
Protection Class
Class 1
Protection Against Ingress of Water
Ordinary (IPXO)
Mode of Operation
Continuous
I/O Connections
Keyboard DIN-5
Type
Barcode
D-SUB 9 Type