Day-to-Day Operation
4
STERRAD
®
200 User’s Guide
51
Watching a Cycle
You can monitor the progress of a cycle by watching the display; it indicates the
phase of the cycle and certain process parameters. A long beep signals that the
cycle is complete.
The display indicates the status of the unit at all times: the current stage of the
sterilization cycle; the temperature in the chamber; the pressure in the chamber;
the cycle start time, and the cycle estimated end time. You can change the display
to show a graph or a numerical readout of the cycle. Each load goes through nine
consecutive stages: vacuum, pre-plasma, injection, diffusion, plasma, injection,
diffusion, plasma, and vent.
Figure 18. Cycle In Progress–Graph Display. Your display may look slightly different.