84
7
3. To help you see the test strip port in reduced lighting, when
you insert a test strip into the PDM, you can turn on the port
light. Press the middle
Soft Key
labeled
Light
. To turn the light
off, press
Light
again.
4. Wait for the PDM to display “Apply a blood sample to the
strip” on the screen (Figure 7-5).
5. Remove the vial cap and place it on a flat surface with
the top of the cap pointing up. Squeeze the vial to dis-
card the first drop. Then, wipe both the tip of the control
solution vial and the top of the cap with a clean, damp
tissue or cloth. Next, squeeze a drop into the small well
on the top of the cap or onto another clean, non-absor-
bent surface.
6. Gently touch only ONE SIDE of the test strip to the control
solution drop. Specifically, apply the control solution to the
test strip side next to the gold metallic channel (Figure 7-6 on
the previous page). If
BG sound
is set to
On
, the PDM beeps
when the test strip is full.
If you do not apply the sample within 10 seconds, the strip
port light turns off. To turn it on again, press
Light
.
The Countdown Screen appears while the sample is process-
ing (Figure 7-7). The strip port light turns off when the sample
is accepted.
Figure 7-5
Figure 7-6
Do not apply control solution to both sides of the test
strip.
Figure 7-7
Figure 7-8