Understanding and Adjusting Basal Rates
20
3
3. Choose the program you want from the list (Figure 3-7), then
press
Enable
.
4. Press
Enable
again to start the selected basal program. If a
Pod is active, the PDM beeps to indicate that the chosen
basal program is running.
Change, rename, copy, or delete a basal program
1. On the
Home
screen, use the
Up/Down Controller
buttons
to choose
Settings
; then press
Select
.
2. Choose
Basal programs
, then press
Select
.
3. Choose a program from the list, then press
Select
.
A diamond icon (
u
) appears next to the program that
is active. You cannot select the active program to
enable it, because it is already running.
Figure 3-7
As a safety feature, you cannot enable a new basal
program while a temporary basal program is in pro-
cess (see “Temporary Basal Rates and Presets” later in
this chapter); you must first cancel the active tempo-
rary basal program.