10
Temperature too
high. Out of operating
range.
For customer support
call 800.591.3455.
•
Meter is too hot (above
111°F) to work correctly.
1. Move the meter and test strips to a cooler area. Insert a new test strip
when the meter and test strips are within the operating range (43–
111°F). If you do not get another Temperature too high message, you
can proceed with testing.
Temperature too low.
Out of operating
range.
For customer support
call 800.591.3455.
•
Meter is too cold (below
43°F)to work correctly.
1. Move the meter and test strips to a warmer area. Insert a new test
strip when the meter and test strips are within the operating range
(43–111°F). If you do not get another Temperature too low message,
you can proceed with testing.
Alert error message
Possible causes
Action