![Insulet mylife OmniPod Manual Download Page 79](http://html1.mh-extra.com/html/insulet/mylife-omnipod/mylife-omnipod_manual_2070293079.webp)
Using the Personal Diabetes Manager
67
6
buttons to choose a new start time, end time, and IC ratio.
Press
Next
after each entry.
b. To add a segment, choose
[add new]
, then press
New
.
Use the
Up/Down Controller
buttons to choose a start
time, end time, and IC ratio. Press
Next
after each entry.
c. When you have completed all time segments, press
Done
, then
Save
.
4.
Correction factor
—You can edit existing time segments and
add segments, up to a total of 8.
a. To edit an existing correction factor time segment, choose
the segment, then press
Edit
. Use the
Up/Down Control-
ler
buttons to choose a new start time, end time, and cor-
rection factor. Press
Next
after each entry.
b. To add a segment, choose
[add new]
, then press
New
.
Use the
Up/Down Controller
buttons to choose a start
time, end time, and correction factor. Press
Next
after
each entry.
c. When you have completed all time segments, press
Done
, then
Save
.
5.
Reverse correction
—Choose
On
or
Off
, then press
Select
.
6.
Insulin action (duration) or Insulin on Board (IOB)
—Use
the
Up/Down Controller
buttons to choose a duration of
insulin action, then press
Enter
.
Temp basal
Choose
%
,
U/hr
, or
Off
to set the mode for temporary basal rates
and presets, then press
Select
.
Extended
Choose
%
,
Units
, or
Off
to set the mode for extended boluses,
then press
Select
.
Bolus increment
Choose
0.05
,
0.10
,
0.50
, or
1.00
units for bolus increments, then
press
Select
.
Max bolus
Use the
Up/Down Controller
buttons to enter the maximum
bolus dose you can take, then press
Enter
.
Max basal rate
Use the
Up/Down Controller
buttons to enter the maximum
basal rate you can choose, then press
Enter
.
Some settings have preset defaults, but all settings
are changeable. See the Appendix for a list of Sys-
tem specifications, including preset default settings.
For additional details on bolus and basal settings,
see Chapter 3, Understanding and Adjusting Basal
Rates, and Chapter 4, Understanding and Deliver-
ing Bolus Doses.
Summary of Contents for mylife OmniPod