160
Radiated RF
(IEC 61000-4-3)
80 MHz–2.5 GHz
10 V/m
Portable and mobile RF communications equipment should
be used no closer to any part of the System than the recom-
mended separation distance calculated from the equation
applicable to the frequency of the transmitter as below.
Recommended separation distance:
d=1.17
P 150 KHz to 80 MHz
d=0.35
P 80 MHz to 800 MHz
d=0.7
P 800 MHz to 2.5 GHz
where P is the maximum output power rating of the transmit-
ter in watts (W) according to the transmitter manufacturer and
d is the recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey,
a
should be less than the compli-
ance level in each frequency range.
Interference may occur in the vicinity of equipment marked
with the following symbol:
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflections from struc-
tures, objects, and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio,
AM and FM radio broadcast, and TV broadcast, cannot be predicted theoretically with accuracy. To assess the electromagnetic environment
due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the
equipment is used exceeds the applicable RF compliance level above, the equipment should be observed to verify normal operation. If
abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the equipment.
Electromagnetic Immunity (Continued)