159
Electromagnetic Immunity
The System is intended for use in the electromagnetic environment specified below. You should observe these requirements in the
use of the System.
Immunity against
IEC 60601-1-2
test level
Compliance level
(of this device)
Electromagnetic environment
ElectroStatic Discharge,
ESD (IEC 61000-4-2)
contact discharge:
± 6 kV
air discharge: ± 8 kV
± 8 kV
± 15 kV
If floors are covered with synthetic material, try to avoid electro-
static discharges.
Power Frequency mag-
netic fields 50/60 Hz
(IEC 61000-4-8)
3 A/m
400 A/m
Suitable for most environments. Magnetic field strengths in
excess of 400 A/m would be unlikely except in close proximity
to industrial magnetic devices.