Pod expiration advisory.
Change Pod now.
Advisory
Alarm
Pod
and
PDM
2 sets of beeps repeat periodically.
Repeats every 15 minutes during
the last hour of the Pod’s life and is
independent of the Expiration noti-
fication that you can set. (See Using
the Personal Diabetes Manager in
Chapter 6.)
Escalates to
Pod expired
hazard
alarm.
Press
OK
to acknowledge.
Change Pod. (See Chapter 5, Using the
Pod.)
Pod expired.
Insulin delivery stopped.
Change Pod now.
Hazard
Alarm
Pod
and
PDM
Continuous tone.
This alarm occurs when the Pod has
reached the end of its operating life.
Press
OK
to acknowledge.
Change Pod.
Check blood glucose.
Stuck key detected.
Check
Home
button to
resolve.
For technical support refer
to your
User Guide
.
Advisory
Alarm
PDM
2 sets of beeps every 5 minutes.
Check the key described on the screen.
(In this example, the stuck key is the
Home
button.) Press the key to unstick
it.
If this does not work, notify Customer
Care.
Alarm
Type
Source
Beep Pattern
Action