129
10
Auto-off.
Remove Pod now.
Hazard
Alarm
Pod
and
PDM
Continuous tone.
Pod has been deactivated. Insulin deliv-
ery has stopped. Press
OK
to acknowl-
edge.
Change Pod.
Check blood glucose.
Low reservoir.
Replace Pod soon.
(This is based on the vol-
ume level you selected.)
Advisory
Alarm
Pod
and
PDM
2 sets of beeps every minute for 3
minutes.
Repeats every 60 minutes until
acknowledged.
Escalates to
Empty reservoir
hazard alarm when empty.
Press
OK
to acknowledge.
Empty reservoir.
Insulin delivery stopped.
Change Pod now.
Hazard
Alarm
Pod
and
PDM
Continuous tone.
This alarm occurs when the Pod's
reservoir is out of insulin. Press
OK
to
acknowledge.
Change Pod.
Check blood glucose.
Alarm
Type
Source
Beep Pattern
Action
The
Low reservoir
alert will
escalate to an
Empty reservoir
hazard alarm when insulin is
depleted. Be sure to respond to
the alert when it first occurs.