140
11
At either “Pod status not available” screen, press
Home
to return
to the
Home
screen or press
Status
to send another status
request.
n
Failure During Bolus Cancellation
When you instruct the PDM to cancel a bolus, if the PDM cannot
send your instruction to the Pod within a few seconds, it displays
the “Communication error” screen (Figure 11-12).
1. Move the PDM close to the Pod. The PDM tries to communi-
cate with the Pod again.
2. If the PDM still cannot communicate with the Pod, you will
see the “Bolus status unknown” message (Figure 11-13).
If the Pod has expired, you will see the screen in
“” on page 140 instead.
Figure 11-11
Figure 11-12
Figure 11-13