152
n
Pod Specifications
Size:
3.9 cm x 5.2 cm x 1.45 cm
Weight (without insulin):
25 grams
Operating temperature range:
4.4°C to 40°C
Under normal circumstances, your body temperature will keep
the Pod well within this range.
Startup temperature:
above 10°C
Storage temperature range:
0°C to 30°C
Reservoir volume (deliverable):
200 units
Cannula insertion depth:
6.5 mm
Waterproof rating:
IPX8 (7.6 meters for up to 60 minutes)
Insulin concentration:
U-100
Alarm type:
Audible
Operating relative humidity range:
20–85%, non-condensing
Storage relative humidity range:
20–85%, non-condensing
Operating atmospheric pressure:
696 hPA to 1060 hPA
Storage atmospheric pressure:
696 hPA to 1060 hPA
Non-pyrogenic:
Fluid pathway only
Type BF Medical Device:
Protection from electrical shock
Flow Capability:
Basal: Programmable by the user in 0.05 U pulses
up to 30.0 U per hour
Bolus:
Rate: 1.5 units per minute
Range: 0.05 to 30.0 units
Delivery accuracy (tested per IEC 60601-2-24):
Basal: +/- 5% at rates > 0.05 U/hr
Bolus: +/- 5% for all set values > 0.05 units
Bolus presets
7
Carbohydrate presets
36
Suspend
30 minutes to 2 hours
Low reservoir volume
indicator
10 to 50 units in 5-unit increments
Default is 10.0 units
Pod expiration alert
1 to 24 hours in 1-hour increments
Default is 4 hours
Auto-off alert
1 to 24 hours or Off, in 1-hour
increments
Default is Off
PDM lock
On or Off
Default is Off
History storage
5400 records/90 days
Language
English