Multitron Standard - Operating Manual
Installation
Page
50
of
114
17 March 2021
Placing the device under a bench
When placing the device under a bench, observe the following
points:
You must leave a ventilation gap of at least 100 mm between
the device and rear wall of the bench to ensure that heat pro-
duced by the device can escape.
The minimum distance between the device and the underside
of the bench top is 100 mm.
ATTENTION
Failure to comply with the required minimum distances can result
in components of the device overheating and the device being
damaged. As such, please observe the following points:
–
Never cover the air vents on the right-hand side or those on
the rear of the device.
–
Never position the device right up against a wall.