Multitron Standard - Operating Manual
Operation
17 March 2021
Page
55
of
114
ATTENTION
If the tray is not held in place securely on the table, the tray may
be damaged by the movement of the table inside the machine.
–
Do not start up the device unless the tray has been inserted
correctly and is held in place securely by the lock hooks.
INFORMATION
If you are unable to insert the tray properly, or the system does not
lock it in place securely, perform the following checks:
–
Remove the tray and check it is not twisted or bent.
–
Check the incubation chamber for foreign matter and other
objects.
–
If it is still not possible to insert the tray correctly, contact
your nearest INFORS HT representative.
6.2.4
Handling Cultivation Flasks without Removing the Tray
The cultivation flasks can also be handled without removing the tray.
To do so, proceed as follows:
CAUTION
If the door is opened when the shaker drive is running, the table
can cause injury due to its significant momentum. As such, take
care when opening the door while the shaker drive is running, and
wait until the table has come to a complete standstill.
Procedure
1.
Temporarily stop the device by slightly opening the door.
2.
Wait until the table has come to a standstill.
3.
Open the device door max. 80°. If the door is opened more than
80°, the tray will be released.
4.
Handle the cultivation flasks.
5.
Close the door again.
The device restarts automatically.