Multitron Standard - Operating Manual
Maintenance and Cleaning
Page
92
of
114
17 March 2021
8
Maintenance and Cleaning
WARNING
To prevent life-threatening electric shocks, always switch off the
device and disconnect it from the power supply before carrying
out any maintenance or cleaning.
8.1
Maintenance
The device requires hardly any maintenance. This reduces the running
costs to certain regular checks and cleaning. The following table de-
scribes the maintenance work that is required to ensure optimum,
fault-free operation.
If increased wear is detected during the regular checks, the required
maintenance intervals must be reduced to suit with the actual signs
of wear. Keep in mind that various media or gases will have more or
less corrosive effects on the metal parts. When using particularly ag-
gressive substances, more frequent checks are required in order to
maintain smooth device operation.
Contact the manufacturer if you have any questions regarding
maintenance work or intervals; see Service contact details on page 2.
Interval
Maintenance work
To be carried out by
Prior to each use Check all seals on the device, especially on the door, and replace
if necessary.
Qualified technicians
Check whether the interior lighting works; have lamps replaced if
necessary.
Qualified technicians
After each use
Clean the device; if necessary, disinfect thoroughly.
Operator
Annually
Have sensors calibrated at least once per year to ensure that the
measuring results remain accurate.
Qualified technicians
Every 3 years
If using the cooling option, have the coolant replaced. To do this,
contact INFORS HT.
INFORS HT service tech-
nician or licensed dealer