Splashing water
Always use a wet cloth when cleaning the base tray; never pour wa-
ter into the tray. Make sure that no water splashes into the bearings.
After cleaning the device, especially the interior and the base tray, dry
it with a cloth.
Dust
Over time, dust can collect in the vents and on other exposed parts. It
can be removed carefully using a vacuum cleaner.
Installing/uninstalling the table
In order to allow cleaning of the base tray in case of glass breakages
or soiling caused by culture liquids, the table can be detached from
the counterweight and folded up. To do so, proceed as follows:
ATTENTION
After being folded up, the table must be held in position by hand.
If the table is not secured properly, there is a risk of it folding back
down again. This can cause injuries.
–
If possible, have a second person hold the table in place once
it has been folded up. Alternatively, you can place an object
under the table to stop it from folding back down.
–
Proceed with caution when folding the table back down.
Procedure
1.
Open the device door all the way.
2.
Remove the tray.
3.
Turn off the power switch and pull out the plug.
4.
Undo the hexagonal socket screws (4 M5x16) on the table’s
drive hub using the supplied Allen key and remove them.
ATTENTION
No screws must be removed other than the four mounting screws.
5.
Tip the table up by a maximum of 30°.
6.
Hold the table in its folded-up position (or, ideally, get a second
person to do so), or use a suitable object to hold it there.