Multitron Standard - Operating Manual
Technical Data and Specifications
Page
106
of
114
17 March 2021
11.3.2
Temperature Parameter
Description
Value Unit
Ventilation
2 heated cross-flow fans
Power consumption
750 W
Air circulation
360 m
3
/h
Lowest temperature
(via ambient temperature, without optional
cooling)
6
°C
Highest temperature
65 °C
Control precision 4 – 50°C
± 0.3 °C
Control precision > 50°C
± 0.5 °C
Control system
Electronic PID control system
Sensor type
Pt100
Measurement range of the sensor
4 – 80 °C
Capillary thermostat control preci-
sion
± 2.5 °C
Default thermo switch setting for
capillary thermostat
70 °C
Bi-metallic heating switch
approx. 80 °C
Overheating protection (safety
fuse)
84 °C
Coolant for optional cooling system R134a (non-CFC)
11.4
Parameter Limits
Depending on the options installed, the laws of physics may limit the
parameter setpoints in terms of what can actually be achieved.
All this information is non-binding and merely provided as an over-
view; the values that can be achieved in practice depend on many
factors, which cannot all be taken into account here.