Multitron Standard - Operating Manual
Operation
Page
52
of
114
17 March 2021
CAUTION
If you put a load of more than 20 kg on the door, there is a risk
that the device might fall down. This can lead to property damage
and injuries.
–
Do not lean on the door.
–
Do not put a load of more than 20 kg on the door.
6.2.1
Opening the Door
The door can be opened when the device is running. The table will
come to a stop as soon as the door is opened more than 30°. It is not
possible to open the door fully until the table has come to a full stop.
The table will start up again automatically as soon as the door is
closed more than 45°.
CAUTION
If the door is opened when the shaker drive is running, the table
can cause injury due to its significant momentum. As such, take
care when opening the door while the shaker drive is running, and
wait until the table has come to a complete standstill.
ATTENTION
The door can only be opened fully when the device is switched on
at the power switch and there is no load on the half-open door. If
the door is pushed down violently, this can damage the door
mechanism.
In order to open the door while the shaker drive is running, proceed
as follows:
Procedure
1.
Temporarily stop the device by slightly opening the door.
2.
Wait until the table has come to a standstill.
3.
Open the device door all the way. If the door does not unlock
straight away, do not lean on it; instead, lift it up slightly to take
the load off the locking mechanism.