INFICON
Service Instructions | 5
Sensistor Sentrac-Leak Detector-ServiceManual-nip65en1-05-(1901)
43 / 76
ð
Blue LED is lit.
ð
Display is lit up and functions and no “rainbow rings” are visible.
ð
Touch function on the display is working.
ð
Sound in earphone and speaker is working.
ð
Probe is working in both front and rear Probe sockets.
ð
Electrical grounding is OK. Measure with a DVM between front probe
connection ground (the shield) and one of the screws in the back. The
resistance should be 9 ohm or lower.
11
If any of the above tests fails, see Troubleshooting [
21].