5.19.2 Changing Probe neck O-ring
Part
Part no.
Qty
Consumables
O-ring 8x1 NBR70
591-284
1
Disassemble
1
Disconnect the Probe from the Sensistor Sentrac.
2
Use a Sensor Key to remove the Probe Sensor Nut (A) from the Probe Neck
(B).
3
Remove the Probe Neck O-ring (C) from the Probe Neck (B).
Assemble
1
Mount the new Probe Neck O-ring (C) onto the Probe Neck (B).
2
Use a Sensor Key to mount the Probe Sensor Nut (A).
3
Connect the Probe to the Sensistor Sentrac.
4
Start the Probe and make sure that it works properly, check:
ð
Sensitivity
ð
Display
ð
LED-lamp
5
If the test fails, see Troubleshooting [
21].