5 | Service Instructions
INFICON
50 / 76
Sensistor Sentrac-Leak Detector-ServiceManual-nip65en1-05-(1901)
4
Measure the resistance between the Probe tip and the C21 connection port.
Resistance should be lower than 9 ohm. Re-tension the C21 connection port nut
if necessary.
5
Connect the Probe to the Sensistor Sentrac.
6
Start the Probe and make sure that it works properly, check:
ð
Sensitivity
ð
Display
ð
LED-lamp
7
If the test fails, see Troubleshooting [
21].