INFICON
Service Instructions | 5
Sensistor Sentrac-Leak Detector-ServiceManual-nip65en1-05-(1901)
57 / 76
5.19 Changing Probe O-rings
5.19.1 Changing Nut O-ring
Part
Part no.
Qty
Consumables
O-ring 8x1 NBR70
591-284
1
Disassemble
1
Disconnect the Probe from the Sensistor Sentrac.
2
Use a Sensor Key to remove the Probe Sensor Nut (A) from the Probe Neck
(B).
3
Use a small slotted screw driver to remove the O-ring inside the Probe Sensor
Nut (A).
4
Clean the Probe Sensor Nut (A) thoroughly.
Assemble
1
Use the O-ring mounting tool for nut probe tip to mount the O-ring into the Probe
Sensor Nut (A).
2
Lubricate the O-ring with lock oil.
3
Use a Sensor Key to mount the Probe Sensor Nut (A).
4
Connect the Probe to the Sensistor Sentrac.
5
Start the Probe and make sure that it works properly, check:
ð
Sensitivity
ð
Display
ð
LED-lamp
6
If the test fails, see Troubleshooting [
Summary of Contents for Sensistor Sentrac
Page 2: ...INFICON AB Westmansgatan 47F S 58216 Lingköping Sweden ...
Page 9: ...INFICON Components 3 Sensistor Sentrac Leak Detector ServiceManual nip65en1 05 1901 9 76 ...
Page 10: ...3 Components INFICON 10 76 Sensistor Sentrac Leak Detector ServiceManual nip65en1 05 1901 ...
Page 13: ...INFICON Components 3 Sensistor Sentrac Leak Detector ServiceManual nip65en1 05 1901 13 76 ...
Page 14: ...3 Components INFICON 14 76 Sensistor Sentrac Leak Detector ServiceManual nip65en1 05 1901 ...
Page 72: ...7 Drawings INFICON 72 76 Sensistor Sentrac Leak Detector ServiceManual nip65en1 05 1901 ...
Page 73: ...INFICON Drawings 7 Sensistor Sentrac Leak Detector ServiceManual nip65en1 05 1901 73 76 ...
Page 74: ...Index INFICON 74 76 Sensistor Sentrac Leak Detector ServiceManual nip65en1 05 1901 Index ...
Page 75: ......
Page 76: ......