INFICON
Service Instructions | 5
Sensistor Sentrac-Leak Detector-ServiceManual-nip65en1-05-(1901)
31 / 76
5.5 Changing Power Supply (Desktop Model)
Part
Part no.
Qty
Consumables
Power supply
RS-75-24
598-195
1
Disassemble
1
Disconnect the mains connector.
2
Remove the cover, see Disassemble in Remove the Cover [
28].
3
Remove the two screws (A) holding the Power Supply (B).
4
Disconnect the power supply cabling.
5
Remove the Power Supply (B).
Assemble
1
Install the new Power Supply.
2
Connect the power supply cabling according to the drawing below.
3
Reinstall and tighten the two screws, 4 mm, (A) holding the Power Supply.