6 | Software
INFICON
64 / 76
Sensistor Sentrac-Leak Detector-ServiceManual-nip65en1-05-(1901)
6.3 Service Screen
This part contains information regarding tools in order to analyze the function and
identify equipment failures.
1
Click Diagnosis >> Service Screen.
2
Fill in the Service Password.
6.3.1 Signals
• Sensor Voltage - Raw signal from the sensor in mV.
• Measuring Value - Same Measuring value is displayed in Measuring Mode.
• Locating Value - The numeral is a unitless value of the output signal from the
locating algorithm. The percent describe the bar length on the screen.
• Sensitivity - The senor sensitivity. For 10 ppm H2 the value can be between 0.1 to
5 mV/s, depending of operating time and gas exposure history.
• Power - The operating voltage (V).
• Sensor Diode - Indicates the sensor temperature by the voltage drop over a diode.
Normal voltage drop: 600 ± 20 mV.
• Conversion Factor - The internal unit and gas conversions parameter between
calibration settings and Measuring Mode settings.
• If unit and gas settings are equal, the Conversion Value = 1.
• Calibration Factor - Internal calibration factor.
6.3.2 Locating Graph
Displays the Locating Value graphically.
6.3.3 Debug Settings
Print out of the following parameters on RS232 or USB:
• Time counter in ms (milli seconds)
• A/D value (sensor signal) raw (24 bit) or in mV (depending on settings)
• Locating Value
• Measuring Value in mV/s (uncalibrated and unconverted)
• Data Checksum
Output frequency 1-50 Hz.
Summary of Contents for Sensistor Sentrac
Page 2: ...INFICON AB Westmansgatan 47F S 58216 Lingköping Sweden ...
Page 9: ...INFICON Components 3 Sensistor Sentrac Leak Detector ServiceManual nip65en1 05 1901 9 76 ...
Page 10: ...3 Components INFICON 10 76 Sensistor Sentrac Leak Detector ServiceManual nip65en1 05 1901 ...
Page 13: ...INFICON Components 3 Sensistor Sentrac Leak Detector ServiceManual nip65en1 05 1901 13 76 ...
Page 14: ...3 Components INFICON 14 76 Sensistor Sentrac Leak Detector ServiceManual nip65en1 05 1901 ...
Page 72: ...7 Drawings INFICON 72 76 Sensistor Sentrac Leak Detector ServiceManual nip65en1 05 1901 ...
Page 73: ...INFICON Drawings 7 Sensistor Sentrac Leak Detector ServiceManual nip65en1 05 1901 73 76 ...
Page 74: ...Index INFICON 74 76 Sensistor Sentrac Leak Detector ServiceManual nip65en1 05 1901 Index ...
Page 75: ......
Page 76: ......