5 | Service Instructions
INFICON
60 / 76
Sensistor Sentrac-Leak Detector-ServiceManual-nip65en1-05-(1901)
5.21 Functional Verification
5.21.1 Functional Verification of Sensitor Sentrac
1
Connect the Hand Probe P60 to the front connection port on the Sensistor
Sentrac.
2
Turn the Sensistor Sentrac unit on.
3
Verify that the Hand Probe P60 has the same software version number as
stated in the Sensistor Sentrac Software Installer.
If not: update the probe via USB cable.
4
Desktop Model: connect the power cable and verify that the plug is properly
mounted.
5
Verify that the blue LED on the panel lights up and pulsates.
6
Wait for warm-up to be completed.
7
If the display is changed, calibrate the touch-screen according to the instructions
on the display. Verify that:
ð
Display
ð
no staining (rainbow colors) is displayed. If so, the display is positioned
incorrectly and in tension.
ð
is well centered.
ð
contrast and brightness is good and the picture is stable (no flicker).
adjustment of brightness.
ð
no pixel / graphical glitches.
ð
touch function is operational.
ð
Grounding from the AC socket and ground exist. Use a multimeter to
measure between the center pin on the AC socket and the edge of the probe
cable connector. The result should be 9 ohm or lower.
ð
Rear connectors are well centered in their position holes.
ð
Front panel LEMO connector is fastened properly and turned in the right
direction (red dot facing up).
ð
Rear panel LEMO connector is fastened properly.
8
Function control of rear outputs:
ð
Device Port (probe connector)
ð
Connect Hand Probe P60 to Device Port.
ð
Turn the Sensistor Sentrac unit on.
ð
Verify that the probe and instrument function together.
ð
Leave the Sensistor Sentrac unit on.
Summary of Contents for Sensistor Sentrac
Page 2: ...INFICON AB Westmansgatan 47F S 58216 Lingköping Sweden ...
Page 9: ...INFICON Components 3 Sensistor Sentrac Leak Detector ServiceManual nip65en1 05 1901 9 76 ...
Page 10: ...3 Components INFICON 10 76 Sensistor Sentrac Leak Detector ServiceManual nip65en1 05 1901 ...
Page 13: ...INFICON Components 3 Sensistor Sentrac Leak Detector ServiceManual nip65en1 05 1901 13 76 ...
Page 14: ...3 Components INFICON 14 76 Sensistor Sentrac Leak Detector ServiceManual nip65en1 05 1901 ...
Page 72: ...7 Drawings INFICON 72 76 Sensistor Sentrac Leak Detector ServiceManual nip65en1 05 1901 ...
Page 73: ...INFICON Drawings 7 Sensistor Sentrac Leak Detector ServiceManual nip65en1 05 1901 73 76 ...
Page 74: ...Index INFICON 74 76 Sensistor Sentrac Leak Detector ServiceManual nip65en1 05 1901 Index ...
Page 75: ......
Page 76: ......