INFICON
Troubleshooting | 4
Sensistor Sentrac-Leak Detector-ServiceManual-nip65en1-05-(1901)
23 / 76
Message
Indicators
Technical
Description
Cause of Warning
Measure
Error Code
Sensor heat
regulation error.
Change sensor
or probe.
Red measure
window present
on screen. Blue
LED flashing.
Sensor
temperature out
of range.
Temperature
check wire (Pin 2)
has voltage < 545
mV. Cause of
failure: Broken
sensor, cable short
circuit
(Temperature
check wire, Pin 2)
or hardware error.
Temperature
check wire (Pin 2)
has voltage > 621
mV. Cause of
failure: Ground
wire broken (Pin
1), Temperature
check wire (Pin 2)
broken, Heating
wire (Pin 4, 5)
broken, sensor
broken or
hardware error.
Change sensor or
probe. If a Combox
is used, change
the cable between
the Combox and
the probe.
03
Probe
disconnected.
Reconnect the
probe.
Red measure
window present
on screen, Blue
LED flashing.
The probe has
been
disconnected.
No cable or broken
cable. Ground wire
(Pin 1) and/or
Probe sens wire
(Pin 5) broken.
Reconnect the
probe or change
the cable. If the
problem remains,
change to new
probe.
04
Probe
communication
problem. Try to
restart the
instrument or
change the
probe cable. If
the problem
remains contact
service.
Red measure
window present
on screen, Blue
LED flashing.
Broken
connection
between probe
and instrument.
Broken cable.
Probe CANH (Pin
2) and/or Probe
CANL (Pin 3) wire
broken.
Change probe
cable or probe.
05
Summary of Contents for Sensistor Sentrac
Page 2: ...INFICON AB Westmansgatan 47F S 58216 Lingköping Sweden ...
Page 9: ...INFICON Components 3 Sensistor Sentrac Leak Detector ServiceManual nip65en1 05 1901 9 76 ...
Page 10: ...3 Components INFICON 10 76 Sensistor Sentrac Leak Detector ServiceManual nip65en1 05 1901 ...
Page 13: ...INFICON Components 3 Sensistor Sentrac Leak Detector ServiceManual nip65en1 05 1901 13 76 ...
Page 14: ...3 Components INFICON 14 76 Sensistor Sentrac Leak Detector ServiceManual nip65en1 05 1901 ...
Page 72: ...7 Drawings INFICON 72 76 Sensistor Sentrac Leak Detector ServiceManual nip65en1 05 1901 ...
Page 73: ...INFICON Drawings 7 Sensistor Sentrac Leak Detector ServiceManual nip65en1 05 1901 73 76 ...
Page 74: ...Index INFICON 74 76 Sensistor Sentrac Leak Detector ServiceManual nip65en1 05 1901 Index ...
Page 75: ......
Page 76: ......