5.11 Changing Speaker
Part
Part no.
Qty
Consumables
KB1114 Speaker
591-826
1
WARNING
Electrical hazard!
When function test is performed on an open instrument, there is a risk of electrical
shock.
CAUTION
Caution
Be very cautious when removing the locking on connector J2 (FFC cable, LCD data,
71].
Disassemble
1
Disconnect the mains connector.
2
Remove the cover, see Disassemble in Remove the Cover [
3
Remove the four screws (A) holding the Front Panel (B).
4
Remove all cables between the Front Panel (B) and the PCB. Pay special
attention to J2. Carefully tap the screwdriver / locking alternately left/right to
loosen the cable from the connector. Do not use force!
5
Remove the Front Panel.
6
Remove the four screws (C) holding the Speaker (D).
Summary of Contents for Sensistor Sentrac
Page 2: ...INFICON AB Westmansgatan 47F S 58216 Lingköping Sweden ...
Page 9: ...INFICON Components 3 Sensistor Sentrac Leak Detector ServiceManual nip65en1 05 1901 9 76 ...
Page 10: ...3 Components INFICON 10 76 Sensistor Sentrac Leak Detector ServiceManual nip65en1 05 1901 ...
Page 13: ...INFICON Components 3 Sensistor Sentrac Leak Detector ServiceManual nip65en1 05 1901 13 76 ...
Page 14: ...3 Components INFICON 14 76 Sensistor Sentrac Leak Detector ServiceManual nip65en1 05 1901 ...
Page 72: ...7 Drawings INFICON 72 76 Sensistor Sentrac Leak Detector ServiceManual nip65en1 05 1901 ...
Page 73: ...INFICON Drawings 7 Sensistor Sentrac Leak Detector ServiceManual nip65en1 05 1901 73 76 ...
Page 74: ...Index INFICON 74 76 Sensistor Sentrac Leak Detector ServiceManual nip65en1 05 1901 Index ...
Page 75: ......
Page 76: ......