5 | Service Instructions
INFICON
42 / 76
Sensistor Sentrac-Leak Detector-ServiceManual-nip65en1-05-(1901)
4
Remove all cables between the Front Panel (B) and the PCB. Pay special
attention to J2. Carefully tap the screwdriver / locking alternately left/right to
loosen the cable from the connector. Do not use force!
5
Remove the Front Panel.
6
Remove the six screws (C) holding the Display (D) on the Front Panel (B).
7
Remove the display plates (E) by removing the four screws (F).
Make note of the display plate positions. The display plates are different.
8
Remove the Display (D).
Assemble
NOTICE
Depending on Display type, jumpers J20 and J23 on PCB may need to be altered.
See Drawings [
Kyocera type No.: TCG057QVLCZ-C00
Datavision type No.: DTFS057A5SHLX-D00
1
Cut the Conductive profile in the following lengths: 104, 135.5, 52 and 38 mm.
2
Attach the Conductive profile strips onto the Display.
3
Mount the display plates according to previous notes (step 8 in Disassemble).
Do not tighten to firmly.
4
Carefully insert the Display in the Front Panel. Leave a gap of at least 0.5 mm
between the Display and the Front Panel inside the LED lamp, to avoid that the
display glass sticks to the plastic edge where the EMC gasket seal is.
5
Reinstall the display plates (E) by tightening the four screws (F).
6
Reinstall and tighten the six screws (C) holding the Display (D). Do not use
thread locking compound.
7
Place the Front Panel correctly. Make sure that no cables are pinched between
the Front Panel and the chassis.
8
Reinstall and tighten the four screws (A) holding the Front Panel. Do not use
thread locking compound.
9
Connect all cables between the Front Panel and the PCB. Pay special attention
to J2. Do not use force!
10
Perform a function test before the assembly continues.
Ensure the following functions, check that the:
ð
Display is centered in the Front Panel. If not, loosen the screws holding the
display and align it.
Summary of Contents for Sensistor Sentrac
Page 2: ...INFICON AB Westmansgatan 47F S 58216 Lingköping Sweden ...
Page 9: ...INFICON Components 3 Sensistor Sentrac Leak Detector ServiceManual nip65en1 05 1901 9 76 ...
Page 10: ...3 Components INFICON 10 76 Sensistor Sentrac Leak Detector ServiceManual nip65en1 05 1901 ...
Page 13: ...INFICON Components 3 Sensistor Sentrac Leak Detector ServiceManual nip65en1 05 1901 13 76 ...
Page 14: ...3 Components INFICON 14 76 Sensistor Sentrac Leak Detector ServiceManual nip65en1 05 1901 ...
Page 72: ...7 Drawings INFICON 72 76 Sensistor Sentrac Leak Detector ServiceManual nip65en1 05 1901 ...
Page 73: ...INFICON Drawings 7 Sensistor Sentrac Leak Detector ServiceManual nip65en1 05 1901 73 76 ...
Page 74: ...Index INFICON 74 76 Sensistor Sentrac Leak Detector ServiceManual nip65en1 05 1901 Index ...
Page 75: ......
Page 76: ......