4 | Troubleshooting
INFICON
24 / 76
Sensistor Sentrac-Leak Detector-ServiceManual-nip65en1-05-(1901)
Message
Indicators
Technical
Description
Cause of Warning
Measure
Error Code
Probe Software
not compatible
with the
instrument.
Change probe
or upgrade
probe software.
Red measure
window present
on screen, Blue
LED flashing.
The instrument
software is not
compatible with
the probe
software.
Probe Software
not compatible
with the
instrument.
Change probe or
upgrade the probe
software.
06
Two probes
connected.
Disconnect one
probe and
restart the
instrument.
Red measure
window present
on screen, Blue
LED flashing.
Software can not
handle two
probes.
Two probes
connected to the
instrument.
Disconnect one
probe and restart
the instrument.
07
Instrument
hardware error,
service
recommended.
Red measure
window present
on screen, Blue
LED flashing.
Instrument PCB
broken.
Instrument PCB
broken.
Change Main
PCB.
08
Probe hardware
error, service
recommended.
Red measure
window present
on screen, Blue
LED flashing.
Probe PCB
broken.
Probe PCB
broken.
Change Probe
PCB.
09
Summary of Contents for Sensistor Sentrac
Page 2: ...INFICON AB Westmansgatan 47F S 58216 Lingköping Sweden ...
Page 9: ...INFICON Components 3 Sensistor Sentrac Leak Detector ServiceManual nip65en1 05 1901 9 76 ...
Page 10: ...3 Components INFICON 10 76 Sensistor Sentrac Leak Detector ServiceManual nip65en1 05 1901 ...
Page 13: ...INFICON Components 3 Sensistor Sentrac Leak Detector ServiceManual nip65en1 05 1901 13 76 ...
Page 14: ...3 Components INFICON 14 76 Sensistor Sentrac Leak Detector ServiceManual nip65en1 05 1901 ...
Page 72: ...7 Drawings INFICON 72 76 Sensistor Sentrac Leak Detector ServiceManual nip65en1 05 1901 ...
Page 73: ...INFICON Drawings 7 Sensistor Sentrac Leak Detector ServiceManual nip65en1 05 1901 73 76 ...
Page 74: ...Index INFICON 74 76 Sensistor Sentrac Leak Detector ServiceManual nip65en1 05 1901 Index ...
Page 75: ......
Page 76: ......