5 | Service Instructions
INFICON
52 / 76
Sensistor Sentrac-Leak Detector-ServiceManual-nip65en1-05-(1901)
5.15 Changing Probe P60 Display Glass
Part
Part no.
Qty
Consumables
Display glass 2.0
OLED
598-242
1
Before dec 2016
Display glass 2.0
OLED
598-259
1
From dec 2017
Adhesive display
glass P60
598-257
1
Before dec 2016
Adhesive display
glass P60
598-260
1
From dec 2017
Disassemble
1
Disconnect the Probe from the Sensistor Sentrac.
2
Open the Probe Cover, see Disassemble in Open Probe P60 Cover [
49].
3
Remove all cables on the Probe PCB (A).
4
Remove the screw (B) holding the Probe PCB (A).
Summary of Contents for Sensistor Sentrac