5.13 Open Probe P60 Cover
Part
Part no.
Qty
Consumables
P60 Probe Handle
Lower Part
598-247
1
LOCTITE
®
2400
Disassemble
1
Disconnect the Probe from the Sensistor Sentrac.
2
Use a Spanner for notched nut and slightly loosen the C21 connection port nut
(A).
3
Remove the two screws and the lock washers (B).
4
Carefully remove the lower part of the Probe handle (C).
Assemble
1
Place the lower part of the Probe handle (C) correctly. Make sure that the cables
are not pinched between the upper and lower part of the Probe handle and that
the lock washer is on the inside of the Probe handle.
2
Reinstall and tighten the two screws and lock washers (B). Use thread locking
compound.
3
Use a Spanner for notched nut and firmly tighten the C21 connection port nut
(A).
Summary of Contents for Sensistor Sentrac
Page 2: ...INFICON AB Westmansgatan 47F S 58216 Lingköping Sweden ...
Page 9: ...INFICON Components 3 Sensistor Sentrac Leak Detector ServiceManual nip65en1 05 1901 9 76 ...
Page 10: ...3 Components INFICON 10 76 Sensistor Sentrac Leak Detector ServiceManual nip65en1 05 1901 ...
Page 13: ...INFICON Components 3 Sensistor Sentrac Leak Detector ServiceManual nip65en1 05 1901 13 76 ...
Page 14: ...3 Components INFICON 14 76 Sensistor Sentrac Leak Detector ServiceManual nip65en1 05 1901 ...
Page 72: ...7 Drawings INFICON 72 76 Sensistor Sentrac Leak Detector ServiceManual nip65en1 05 1901 ...
Page 73: ...INFICON Drawings 7 Sensistor Sentrac Leak Detector ServiceManual nip65en1 05 1901 73 76 ...
Page 74: ...Index INFICON 74 76 Sensistor Sentrac Leak Detector ServiceManual nip65en1 05 1901 Index ...
Page 75: ......
Page 76: ......