7
Remove the Speaker (D).
Assemble
1
Check the speaker gasket. Replace with new if necessary.
2
Place the new Speaker correctly.
3
Reinstall the four screws (C) holding the Speaker.
4
Place the Front Panel correctly. Make sure that no cables are pinched between
the Front Panel and the chassis.
5
Reinstall and tighten the four screws (A) holding the Front Panel. Do not use
thread locking compound.
6
Connect all cables between the Front Panel and the PCB. Pay special attention
to J2. Do not use force!
7
Perform a function test before the assembly continues.
Ensure the following functions, check that the:
ð
Blue LED is lit.
ð
Display is lit up and is functions.
ð
Touch function on the display is working.
ð
Sound in earphone and speaker is working correctly.
ð
Probe is working in both front and rear Probe sockets.
ð
Electrical grounding is OK. Measure with a DVM between front probe
connection ground (the shield) and one of the screws in the back. The
resistance should be 9 ohm or lower.
8
If any of the above tests fails, see Troubleshooting [
21].
9
Summary of Contents for Sensistor Sentrac
Page 2: ...INFICON AB Westmansgatan 47F S 58216 Lingköping Sweden ...
Page 9: ...INFICON Components 3 Sensistor Sentrac Leak Detector ServiceManual nip65en1 05 1901 9 76 ...
Page 10: ...3 Components INFICON 10 76 Sensistor Sentrac Leak Detector ServiceManual nip65en1 05 1901 ...
Page 13: ...INFICON Components 3 Sensistor Sentrac Leak Detector ServiceManual nip65en1 05 1901 13 76 ...
Page 14: ...3 Components INFICON 14 76 Sensistor Sentrac Leak Detector ServiceManual nip65en1 05 1901 ...
Page 72: ...7 Drawings INFICON 72 76 Sensistor Sentrac Leak Detector ServiceManual nip65en1 05 1901 ...
Page 73: ...INFICON Drawings 7 Sensistor Sentrac Leak Detector ServiceManual nip65en1 05 1901 73 76 ...
Page 74: ...Index INFICON 74 76 Sensistor Sentrac Leak Detector ServiceManual nip65en1 05 1901 Index ...
Page 75: ......
Page 76: ......