P1
C
3000 Micro GC Operating Manual
7
Select the correct module address for the new standard GC module from the
drop-down list (see
) to . Each module address corresponds to a
channel number. Valid 3000 Micro GC module addresses are
1
through
4
.
Figure 8-26 Commission a standard GC module - select module
8
After selecting the module address, click
Restart
to incorporate the changes.
See
.
Figure 8-27 Commission a standard GC module - restart
CAUTION
Do not turn off the power without waiting at least three
minutes. 3000 Micro GC must update configuration files
before power is turned off. If power is turned off too soon,
the files could be corrupted, making 3000 Micro GC
inoperable.
9
Wait at least three minutes before shutting down the instrument. See